8S5D

Cryo-EM structure of Arf1-decorated membrane tubules


ELECTRON MICROSCOPY

Starting Model(s)

Initial Refinement Model(s)
TypeSourceAccession CodeDetails
experimental modelPDB 2KSQ 
Sample
Arf1 coated membrane tubules
Specimen Preparation
Sample Aggregation StateHELICAL ARRAY
Vitrification InstrumentLEICA EM GP
Cryogen NameETHANE
Sample Vitrification Details
3D Reconstruction
Reconstruction MethodHELICAL
Number of Particles27285
Reported Resolution (Å)3.79
Resolution MethodFSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry TypeHELICAL
Axial SymmetryC3
Axial Rise5.821
Angular Rotation20.78
Map-Model Fitting and Refinement
Id1 (2ksq)
Refinement SpaceREAL
Refinement ProtocolOTHER
Refinement Target
Overall B Value
Fitting Procedure
Details
Data Acquisition
Detector TypeFEI FALCON IV (4k x 4k)
Electron Dose (electrons/Å**2)21
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelTFS GLACIOS
Minimum Defocus (nm)600
Maximum Defocus (nm)2800
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2.7
Imaging ModeBRIGHT FIELD
Specimen Holder ModelOTHER
Nominal Magnification92000
Calibrated Magnification
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)200
Imaging Details
EM Software
TaskSoftware PackageVersion
PARTICLE SELECTIONcryoSPARCv4.3.1, v4.4.2
IMAGE ACQUISITIONEPU2.2.0.65REL
CTF CORRECTIONcryoSPARCv4.3.1, v4.4.2
MODEL FITTINGISOLDE
MODEL REFINEMENTPHENIX1.21-5187
MODEL REFINEMENTCoot0.9.7 EL
CLASSIFICATIONcryoSPARCv4.3.1, v4.4.2
RECONSTRUCTIONcryoSPARCv4.3.1, v4.4.2
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION1489051